Product Name Alternative:
IFNA1; IFL; LeIF D; IFN; IFN-Alpha; IFNA13; IFN-A; IFNAP22; IFN-Alpha 1b; Interferon Alpha 1b; Interferon, Leukocytic; IFN, Leukocyte; Interferon alpha-DApplications:
Chemiluminescent immunoassay for Antigen Detection.Assay Principle:
The microplate provided in this kit has been pre-coated with an antibody specific to Interferon Alpha (IFNa) . Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Interferon Alpha (IFNa) . Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Interferon Alpha (IFNa) level in the sample or standard.;Assay Protocol:
1. Prepare all reagents, samples and standards;
2. Add 100µ L standard or sample to each well. Incubate 2 hours at 37° C;
3. Aspirate and add 100µ L prepared Detection Reagent A. Incubate 1 hour at 37° C;
4. Aspirate and wash 3 times;
5. Add 100µ L prepared Detection Reagent B. Incubate 30 minutes at 37° C;
6. Aspirate and wash 5 times;
7. Add 100µ L Substrate Solution. Incubate 10 minutes at 37° C;
8. Read RLU value immediately.Assay Performance Time:
2H, 40minSample Type:
Serum, plasma and other biological fluidsDetection Range:
1.37 - 1000 pg/mLPrecision:
Intra-assay Precision (Precision within an assay) : 3 samples with low, middle and high level Interferon Alpha (IFNa) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays) : 3 samples with low, middle and high level Interferon Alpha (IFNa) were tested on 3 different plates, 8 replicates in each plate.
CV (%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Sensitivity:
The minimum detectable dose of this kit is typically less than 0.52pg/mLStability:
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.Specificity:
This assay has high sensitivity and excellent specificity for detection of Interferon Alpha (IFNa) .
No significant cross-reactivity or interference between Interferon Alpha (IFNa) and analogues was observed.Method:
Double-Antibody SandwichOrganism Species:
Chicken (Gallus)