HORIZONTAL AUTOCLAVE

CAT: 1612-ZHA-100 DDVSize: 1 UnitDry Ice: NoHazardous: No
CAT#:1612-ZHA-100 DDVSize:1 Unit
Selected
AVAILABILITY: InStock
24/48H Stock Items & 2 to 6 Weeks non Stock Items.
Product image 1
1 / 1
Features & Benefits
Vessel design life: Internal tank s30408, jacket Q245R life 10 years (20000 sterilization cycles) Design pressure of inner chamber: -0.1~0.3MPa Jacket design pressure: 0.3MPa Built in steam generator (if any) : 0.3MPa Design temperature: 150 ̊C Maximum operating temperature: 138 ̊C Maximum working pressure: 0.25MPa Opening pressure of inner chamber safety valve: 0.28MPa Opening pressure of jacketed safety valve: 0.28 MPa Inner chamber hydrostatic test pressure: 0.39MPa Jacket test pressure: 0.52 MPa Temperature display accuracy: 0.1 ̊C Pressure display accuracy: 0.1KPa Sterilization temperature control accuracy: ±02 ̊C Temperature uniformity: ±1 ̊C Amplitude and times of vacuum pulsation: Amplitude -80~0KPa, Frequency 1~99 Number of pulsations: Amplitude -80~80KPa, Frequency1 Times of positive 1~3 pressure fluctuation: 1~3 Ultimate vacuum: -97 KPa Vacuum leakage rate: ≤0.13KPa/min Surface temperature: The external surface temperature does not exceed the ambient temperature by 20 ̊C Dryness: The weight gain of fabrics after sterilization and drying does not exceed 0.6%, and that of utensils, tools and instruments after drying does not exceed 0.1%. Width x Height: 610 x 910 Depth: 1815 Volume: 1000L Steam Consumption: 55 Length X Width X Height: 2120 x 1410 x 2100 Weight: 1550kg Power Supply: 220±10% V, 50 Hz